8 core Defibrillator leads -Branch


technical specification
Item | (specification) | ||
conductor | Material | Tinned copper + Kevlar(1c) | Tinned copper +Kevlar (7c) |
specification | 165/0. 05TC + 30/0. 05TSN + 200DKevlar | 30/0. 08TC + 200DKevlar | |
Twisted gauge | 22AWG | 26AWG | |
Complete diameter | 0.80MM | 0.51MM | |
insulation | Material | TPE | PP |
thickness | 0.6 | 0.17 | |
diameter | 2. 0 ± 0.05 | 0.85 ± 0.03 | |
colour | red | 7 different colors | |
Cabling | Material | Tissue paper + 1500Dkevlar | |
Wrap thickness | 0.05MM | ||
Overlap rate | 50% | ||
Jacket | Material | TPU (BASF Elastollan 1185AM) | |
thickness | 1 | ||
diameter | 5.80 ± 0.20 | ||
colour | TBD | ||
Performance requirements: | |||
1. Resistance requirements: 22AWG–53.9Ω MAX 26AWG — 131Ω MAX | |||
2. High voltage resistance: core wire 1 and core wire 2, core wire and sheath resistant to DC4000V / 60S, leakage current 0.008mA / MIN 3, DC500V Insulation above 500M0 (test sample line 1 ~). | |||
4. load bending test: | |||
4.1. After bending 10000 times (+/- 90 “, 30 times / mmn 5000 times, rotating 90” around the axis, hanging weight 500g) , 10% of the strands cannot be broken. | |||
4.2. After bending 100000 times (+/- 90 “, 30 times / mmn 50000 times, rotating 90” around the axis and hanging weight 500g), the core wires are continuous. | |||
5. Appearance Requirements: smooth appearance, no defects, unevenness. |