7-leads Shieled Low Noise ECG Cable EC207S

7 leads Low Noise ECG cable
7 core ECG low noise_Cutting diagram





technical specification

itemDescription
ConductorNo. of cores7 (7C)
MaterialTINNED COPPER WIRE+ARAMID FIBER YARN   
Construction8/0. 05+0. 003
InsulationMaterialHDPE
Nom. Thickness0.20
Nom. 0. D.O. 60+0. 05
ColorWhite
Conducting LayerMaterialTPE
Nom’ O. D.0. 75±0. 05
ColorBlack
Inner ShieldMaterialTinned copper
Specification50 ± 5/0.05
Coverage90%
Inner JacketMaterialPE
Nom. Thickness0.10
Nom. O. D.1. 05+0. 15
Color1. White 2. Black 3. Red 4. Green 5. Brown 6. Orange 7. blue
FillerMaterialN/A
Nom. 0. D.N/A
CablingMaterial1000DKEVLAR*6
SeparatorMaterialPTFE(white)
Nom. O. D.3. 4+0. 15
JacketMaterialTPU 5040
Min. Thickness0.40
Nom. 0. D.4. 40+0. 20
ColorMUNSELL 5PB 5/6

Physical and Electrical Performance

ItemRequirement
Hi—potDC5000V/60S /m (conductor—Jacket)
AC3653V/60S /m (conductor—conductor.) AC3653V/60S /m (conductor—shield.)
Noisc(EC53)<50uV
Capacitanceconductor—conductor/shield< 130PF/M. shield—shield<100PF/M
conductor resistance (@20°C)4 ohm/m Max.
Insulation resistance (@20°C)100MQ/5M. Min.
Cable tensile strength>200N lmin
Cable flex life(5N load,5mm bend,
90 degree each side radius. 60cycles/min)
>30000cycles